WAGNER FORCE TEN Manuel d'utilisateur Page 8

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 11
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 7
UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
Ref. No.
SOP/PD/206_01
STANDARD OPERATING PROCEDURE
FORCE TEN
®
FDX
Valid for:
2 years from
approval
Page 8 of 11
5.4.3. Calibration
Start
Press ON/OFF key to
switch on algometer
Allow warm up time of 2-3 minutes
Obtain standard weights that are
equal to 20%, 40%, 60% 80% and 100%
of the capacity of the gauge
Apply each weight consecutively in
both tension and compression
Device is out of tolerance and
requires auto-calibration
Displayed weight
differs by > +/- 0.3%
of full scale +/- 1
least significant
figure
Repeatedly press UNITS key until an
arrow is pointing at lbf (pounds)
Keep device in a still position
Enter Set-Up menu (see section 5.4.1)
Repeatedly press SCROLL key
until [L CAL] is displayed
2
3
1
Yes
No
Vue de la page 7
1 2 3 4 5 6 7 8 9 10 11

Commentaires sur ces manuels

Pas de commentaire