WAGNER FORCE TEN Spécifications Page 10

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 11
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 9
FORCE TEN
FDX FORCE GAGE
18
FORCE TEN
FDX FORCE GAGE
19
CALIBRATION IMPLEMENTS
F. Implements
Several implements are available for use with the FDX - see Table 2. A hook and
at implement are included with FDX. All implements are attached to the American
Standard 10-32 FDX load shaft thread. Metric thread adapters are available.
F.1 Attaching the FDX Hook and Flat Implement
Calibrate:
Apply full-scale weight matching the display - in lbf.
Push SELECT with weight applied and movement stopped.
FDX displays applied weight indicating successful calibration.
FDX will reject a calibration with weights that are higher or lower than FDX
full-scale capacity. It is possible for FDX to accept a calibration with weights
that are close to, but do not match the full scale capacity of FDX. This
erroneous calibration will give inaccurate readings.
Calibration Accepted
FDX displays the actual applied weight indicating successful calibration. If
uuuuu (under) or nnnnn (over) is displayed the weight is not accepted, or other
problems exist - see Calibration Rejected (below).
Vericationisrecommendedbyapplyingweightstocheckaccuracyat
various points of FDX range.
To exit calibration, push SELECT when FDX full-scale capacity is displayed or at
anytime during or after the ve point weight test.
FDX ashes donE and returns to normal operation.
Auto-Calibration is complete and FDX is ready for use.
Calibration Rejected
If the weight is not accepted, the display shows uuuuu (under) or nnnnn (over).
Verify weight applied matches the FDX capacity.
If the weight used does not match FDX capacity, repeat E.3.1 Auto-Calibration
with an accurate full scale weight.
If uuuuu or nnnnn is displayed again after repeating Auto-Calibration, the FDX
may be defective or damaged. Call Wagner Instruments for instructions.