E. Calibration
FDXshouldbeperiodicallytestedtoverifyitiswithinthespeciedtolerance.
E.1 Procedure Description
• Verication-TestingCalibration - Certied weights are applied in tension
and compression to determine if the FDX can be certied and if the calibration
procedure is necessary.
• Calibration - The FDX has Auto-Calibration enabling the gage to be calibrated
by one full capacity test weight.
E.2 Verication-TestingCalibration
Warm-up FDX for 2-3 minutes after power-on to stabilize the electronics.
To verify, a ve-point tension and compression test is recommended with weights
equal to 20%, 40%, 60%,80%, and 100% of FDX capacity. If the displayed weight
differs more than ±0.3% of full scale ±1 least signicant digit, FDX is out of tolerance.
Accuracy verication requires testing xtures: a test stand capable of supporting
weights equal to the FDX capacity, and xtures for applying test weights in tension
and compression. Testing xtures are available from Wagner Instruments.
E.3 Auto-Calibration (L CAL)
Auto-Calibration is used to return the FDX to its specied accuracy after verication
indicates the FDX is out of tolerance, or if obvious that readings are incorrect.
Certied NIST traceable test weights are recommended, resulting in an NIST
Calibration Certicate. Test weights are available from Wagner Instruments.
Unless FDX is tested with E.2 verication and an NIST Calibration Certicate
issued, FDX is not certied to NIST standards. If certication is not required,
Auto-Calibration is adequate in returning FDX to specied accuracy.
AfterAuto-Calibration,FDXaccuracyshouldbeconrmedusing
E.2Verication.
E.3.1 Auto-Calibration (LCAL) Procedure
Test weights must match the Force Cell Module capacity to be tested.
Set UNITS to lbf - all calibrations are performed in pounds (lbf).
Attach FDX to the calibration test stand, and stop all movement.
Enter the set-up menu:
Turn FDX off.
Push / Hold ZERO; push ON/OFF.
Release ON/OFF, then ZERO, AoFF is displayed.
Gage set-up:
SCROLL to L CAL option.
Push SELECT - FDX displays null.
Push SELECT - FDX displays a force cell capacity after zero ashes.
If the capacity shown is not that of the attached Force Cell Module, SCROLL
to the correct Force Cell Module capacity and proceed.