2. If the FDV is used as an interchangeable system, then auto-
calibration is NOT recommended. When the Force Display Module
is calibrated to one Force Cell Module, other Force Cell Modules
cannot be interchanged or fitted to the Force Display Module.
H.3 Auto-Calibration Procedure
1. Enter the Set-Up Menu
a) Turn off the FDV
b) Press and hold SELECT
c) Press POWER
d) Release SELECT and POWER
NOTE: An "Access to Set-up Menu" label is provided on the back of the
FDV gage for quick reference.
2. Scroll through the menu until reaching “CAL”, press SELECT button to
start the calibration procedure.
3. The display will read “null”, since no weights being hung from the gage.
4. Attach any accessories required for hanging test weights.
5. Press SELECT to accept the zero value - a short pause follows.
6. The gage will display “LXXX”, where “XXX” is the capacity of the load cell.
7. Hang weights equal to the capacity of the gage, which should equal
the “XXX” value.
8. Press SELECT to accept this as the full scale weight; the gage will pause.
NOTE: If the test weight differs by more than 15% from the capacity of
the load cell then the gage will not calibrate; the display will show
“nnnn” if the weight is higher than expected, “uuuu” if the weight is
lower than expected.
9. If the calibration is successful the gage will display the full scale
output of the Force Cell Module. Press SELECT to display “donE” and
SELECT again to return to normal operation.